This article assesses the enhance I-Chip across seven ANDE 6C instruments, demonstrating, based on heatmap consistency, that there are few sensitivity differences between those instruments.
Identical sensitivity and mixture sample sets were provided to six external laboratories to evaluate the recent enhancements to the Rapid DNA technology for crime scene-type samples. Data were returned to the Federal Bureau of Investigation and the National Institute of Standards and Technology for review and analysis. Manual interpretation (modified Rapid DNA analysis) was conducted on all samples to assess the success of genotyping for both full profiles and the 20 Combined DNA Index System (CODIS) core STR loci using 18 total cellular DNA input amounts ranging from 102.4 ng to 0.025 ng. Two-person and three-person mixtures at varying ratios for two total DNA inputs were analyzed and observed allele percentages were compared to expected allele percentages. (Published Abstract Provided)
Downloads
Similar Publications
- A quantitative assessment of Saw Kerf floor shape patterns using outline analysis
- DyeSPY: Establishing the First Forensic SERS Reference for Hair Dye Colorant EvidenceArticle link copied!
- An Update on the Presence of BTMPS in the Drug Supply and the Discovery of Tetramethylfentanyl-Related Substances