Each cartridge generated profiles with sufficient data quality to meet CODIS eligibility in fewer than 120 min. The results exhibited 100 percent correlation when compared to conventional DNA typing methods. Precision, reproducibility, stochastic, mixture, and contamination experiments produced expected results. Sensitivity of the ACE sample cartridge was acceptable for buccal swab analysis. The sensitivity of the EXT sample cartridge is discussed. The ACE validation and the EXT evaluation utilized a minimalist, cost-saving, efficient design to generate a validated RapidHIT ID instrument capable of producing genetic profiles from both extracted forensic DNA samples and buccal swab samples within 120 min. (publisher abstract modified)
Downloads
Similar Publications
- Discrimination Between Human and Animal Blood Using Raman Spectroscopy and a Self-Reference Algorithm for Forensic Purposes: Method Expansion and Validation
- Forensic Discrimination of Dyed Hair Color: I. UV-Visible Microspectrophotometry
- IS2aR, a Computational Tool to Transform Voxelized Reference Phantoms into Patient-specific Whole-body Virtual CTs for Peripheral Dose Estimation